20121008 Safety Information
UL leakage current test requirements for new medical ANSI / AAMI standard ES60601-1: 2005

Based on the new edition of medical standards UL ANSI / AAMI ES60601-1: 2005 section 8.7.3 (d) requirements

The leakage current limit value of 5 mA (Normal condition), 10 mA (Single Fault Condition),


Contact current is 100 uA users (Normal condition), 500 uA (Single Fault Condition)

Contrast between old and new standards, we can find relaxation in the leakage current limits in the new

version of many, many manufacturers in designing the product relatively easy, but recent UL inform:

suggest we still use the old version of limit values ​​to request that
the customer

Because in future revisions of the medical standards UL ANSI / AAMI ES60601-1 a high leakage current

may be limited in value back to a more stringent standard in the past. When designing a Medical Equipment

manufacturer with special attention.


For the legacy of medical standards UL limit

Limitations on leakage current value is 300uA (Normal condition); 1000 uA (Single Fault Condition),

Contact current is 100 uA users (Normal condition); 300 uA (Single Fault Condition)